Scanning Digital Monochromator, 420 nm – 1.15 μ
Scanning Digital Mini-Chrom Monochromator Specifications | SDMC1-04V
Operating Range
420 nm – 1.15 μ
f Number
3.9
Focal Length
74 mm
Beam Path
In Line
Resolution with 50μ slits
0.53 nm
Resolution with 100μ slits
1.06 nm
Resolution with 150μ slits
1.59 nm
Resolution with 300μ slits (included)
3.19 nm
Resolution with 600μ slits
6.37 nm
Resolution with 1 mm slits
10.62 nm
Grating Density
1200 grooves/mm
Grating Blaze Wavelength
500 nm
Linear Dispersion (Center of Range)
10.66 nm/mm
Stray Light
≤ 0.003%
Wavelength Accuracy
± 0.2% of λ
Wavelength Reproducibility
± 0.15%
Wavelength Readability
0.2 nm
Operating Temperature Range
-20° C to +80° C
Damage Threshold: Pulsed
350 mJ /cm², 200 nano sec. pulse
Damage Threshold: CW laser
40 Watt/cm²
Optometrics
Scanning Digital Mini-Chrom Monochromator with Operating Range 420 nm - 1.15 μ
Scanning Digital Mini-Chrom Monochromator
The Scanning Digital Mini-Chrom Monochromator (SDMC) includes an integral stepping motor which can be controlled by a calibrated drive such as the Optometrics PCM-02 via a 15-pin connector. The SDMC is similar to the Digital Mini-Chrom in that it includes a four digit counter for visual wavelength readout and a dial for manual wavelength selection.
The wavelength on all digital Mini-Chroms can be selected and read to 0.2 nm. Wavelength is read directly in nanometers (nm) from a counter on models 01, 02, 03, 04 and 05. For near infrared model (06), the four digit counter reading must be doubled, to 2 nm per division.
↓ Scroll down to view product inventory ↓
The Scanning Digital Mini-Chrom can be used for all applications that require the convenience of an integrated visual wavelength readout and manual wavelength tunability, as well as applications requiring computer controlled scanning to a selected wavelength or routine. To prevent mechanical damage due to inadvertently exceeding wavelength limits, all Scanning Digital Mini-Chroms are supplied with dual photosensors that function as high and low limit switches are included for use in the motorized scanning mode.